4-2
4.2 Quick Admitting a Patient
Use [Quick Admit] only if you do not have the time or information to fully admit a patient. Complete the rest of the
patient demographic details later. Otherwise, the symbol will always be displayed in the patient information area.
1. Select the [Patient Setup] QuickKey, or [Main Menu][Patient Setup >>].
2. Select [Quick Admit]. If a patient has been admitted at present, select [Ok] to discharge the current patient. If no
patient is admitted, you can choose either:
[Yes] to apply the data in your monitor to the new patient, or
[No] to clear any previous patient data.
3. Enter the patient category and paced status for the new patient, and then select [Ok].
4.3 Querying and Obtaining Patient Information
The monitor can obtain patient information from HIS through eGateway. To query or obtain patient information from
HIS,
1. Select [Main Menu][Maintenance >>][User Maintenance >>]enter the required password[Gateway
Comm Setting >>], and set [IP Address] and [Port]. Set [ADT Query] to [On].
2. Click the patient information area to enter the [Patient Demographics] menu.
3. Select [Obtain Patient Info. >>] to enter the [Obtain Patient Information] menu.
4. Input the query condition and then select [Query]. The monitor will display the obtained patient information.
5. Select a patient and then click [Import]. The monitor will update the information of the corresponding patient.
6. Select to exit the [Obtain Patient Information] menu.
NOTE
The option [Obtain Patient Information] is available in the [Patient Setup] menu only when [ADT Query] is
set to [On].
When obtaining patient information from HIS, the monitor will only update patient information. The
patients monitoring data is not changed and the patient is not discharged.