5-3
5.6 Deleting a Configuration
1. Select [Delete Config. >>] in the [Manage Configuration] menu.
2. The popup menu shows the existing user configurations on the monitor. Selecting [Config. on USB drive >>] will
show the existing user configurations on the USB drive. Select the user configurations you want to delete and then
select [Delete].
3. Select [Yes] in the popup.
5.7 Transferring a Configuration
When installing several monitors with identical user configurations, it is not necessary to set each unit separately. A USB
drive may be used to transfer the configuration from monitor to monitor.
To export the current monitor’s configuration:
1. Connect the USB drive to the monitor’s USB port.
2. Select [Export Config. >>] in the [Manage Configuration] menu.
3. In the [Export Config.] menu, select the configurations and [User Maintenance Settings] to export. Then select
the [Export] button. A status message will report completion of the transfer.
To import the configuration on the USB drive to the monitor:
1. Connect the USB drive to the monitor’s USB port.
2. Select [Import Config. >>] in the [Manage Configuration] menu.
3. In the [Import Config.] menu, select the configurations and [User Maintenance Settings] to import. Then select
the [Import] button. A status message will report completion of the transfer.
5.8 Loading a Configuration
You may make changes to some settings during operation. However, these changes or the pre-selected configuration
may not be appropriate for the newly admitted patient. Therefore, the monitor allows you to load a desired
configuration so as to ensure that all the settings are appropriate for your patient.
To load a configuration,
1. Select [Load Configuration >>] from the main menu.
2. The popup menu shows the existing configurations on the monitor. Selecting [Config. on USB drive >>] will show
the existing configurations on the USB drive.
3. Select a desired configuration.
4. Select [View] to view the configuration details. In the popup menu, you can select [Alarm Setup >>], [Screen
Setup >>] or [Parameter >>] to view the corresponding contents. The alarm setup items which are different than
those currently used are marked in red.
5. Select [Load] to load this configuration.