15-3
15.2.2 Making a Microstream CO
2
Measurement
Connect the sampling line to the monitor and then connect the CO
2
components as shown below. After warm-up is
finished, you can perform CO
2
measurements
15.3 Changing CO
2
Settings
15.3.1 Accessing CO
2
Menus
By selecting the CO
2
parameter window, you can access the [CO
2
Setup] menu.
15.3.2 Entering the Standby Mode
The standby mode of the CO
2
module relates to the standby mode of the monitor as follows:
If the monitor enters the standby mode, the CO
2
module will also enter the standby mode.
If the monitor exits the standby mode, the CO
2
module will also exit the standby mode.
If the CO
2
module enters or exits the standby mode, it will not affect the monitor.
To enter or exit the standby mode manually, select [Operating Mode] in the [CO
2
Setup] menu and then toggle
between [Standby] and [Measure].
When you set the sidestream CO
2
module to the strandby mode, the CO
2
gas sample intake pump automatically sets the
sample flow rate to zero. When exiting the standby mode, the CO
2
module continues to work at the preset sample flow
rate.
For the sidestream CO
2
module, you can set the delay time. After the delay time the CO
2
module enters the standby
mode if no breath is detected.
Sampling line
Connector for sampling line