E-1
E Electrical Safety Inspection
The following electrical safety tests are recommended as part of a comprehensive preventive maintenance program.
They are a proven means of detecting abnormalities that, if undetected, could prove dangerous to either the patient or
the operator. Additional tests may be required according to local regulations.
All tests can be performed by using commercially available safety analyzer test equipment. These procedures assume
the use of a 601PROXL International Safety Analyzer or equivalent safety analyzer. Other popular testers complying with
IEC 60601-1 used in Europe, such as Fluke, Metron, or Gerb, may require modifications to the procedure. Please follow
the instructions of the analyzer manufacturer.
The electrical safety inspection should be periodically performed every two years .The safety analyzer also proves to be
an excellent troubleshooting tool to detect abnormalities of line voltage and grounding, as well as total current loads.
E.1 Power Cord Plug
Test Item Acceptance Criteria
The power plug
The power plug pins No broken or bent pin. No discolored pins.
The plug body No physical damage to the plug body.
The strain relief
No physical damage to the strain relief. No plug warmth for device in
use.
The power plug No loose connections.
The power cord
No physical damage to the cord. No deterioration to the cord.
For devices with detachable power cords, inspect the connection at
the device.
For devices with non-detachable power cords, inspect the strain relief
at the device.
E.2 Device Enclosure and Accessories
E.2.1 Visual Inspection
Test Item Acceptance Criteria
The enclosure and accessories
No physical damage to the enclosure and accessories.
No physical damage to meters, switches, connectors, etc.
No residue of fluid spillage (e.g., water, coffee, chemicals, etc.).
No loose or missing parts (e.g., knobs, dials, terminals, etc.).

Table of Contents