23-3
23.3 Checking Monitor and Module Information
To view the information about system start time, self-test, etc., select [Main Menu][Maintenance >>][Monitor
Information >>]. You can print out the information for the convenience of troubleshooting. The information will not be
saved during shut down.
You can also view the information about the monitor configuration and system software version by selecting [Main
Menu][Maintenance >>][Software Version >>].
23.4 ECG Verification
The ECG signal may be inaccurate due to hardware or software problems. As a result, the ECG wave amplitude becomes
greater or smaller.
You can print out the square wave and wave scale and then measure the difference between them if necessary. If the
difference exceeds 5%, contact your service personnel.
23.5 NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. It is required at least once a year or when you
doubt the measured NIBP. Contact your service personnel to perform NIBP leakage test.
23.6 NIBP Accuracy Test
The NIBP accuracy test is required at least once a year or when you doubt the measured NIBP. Contact your service
personnel to perform NIBP accuracy test.
23.7 CO
2
Test
For sidestream and microstream CO
2
modules, a calibration is needed every year or when the measured values have a
great deviation. For maintream CO
2
module, no calibration is needed. Contact your service personnel to calibrate CO
2
.