8-5
8.4 Understanding the ECG Display
Your display may be configured to look slightly different.
1. Lead label of the displayed wave
2. ECG gain
3. ECG filter label
4. Notch filter status
Besides, when a paced signal has been detected, the pace pulse marks | are shown on the ECG wave if the [Paced] has
been set to [Yes ].
1. Current heart rate alarm limits
2. Current heart rate
3. Heart beat symbol
8.5 Changing ECG Settings
8.5.1 Accessing ECG Menus
By selecting the ECG parameter window or waveform area, you can access the [ECG Setup] menu.
8.5.2 Choosing the Alarm Source
In most cases the HR and PR numerics are identical. In order to avoid simultaneous alarms on HR and PR, the monitor
uses either HR or PR as its active alarm source. To change the alarm source, select [Alm Source] in the [ECG Setup] menu
and then select either:
[HR]: if you want the HR to be the alarm source for HR/PR.
[PR]: if you want the PR to be the alarm source for HR/PR.