A-1
A Product Specifications
A.1 Monitor Safety Specifications
A.1.1 Classifications
The monitor is classified, according to IEC60601-1:
Type of protection against electrical shock
Class I, equipment energized from an external and internal electrical
power source.
Degree of protection against electrical shock
Type BF defibrillation proof for CO
2
Type CF defibrillation proof for ECG, RESP, TEMP, SpO
2
, N I B P, and IBP
Mode of operation
Continuous
Degree of protection against harmful ingress
of water
IPX1
A.1.2 Environmental Specifications
Main unit
Item Operating conditions Storage conditions
Temperature (℃) 0 to 40 -20 to 60
Relative humidity (noncondensing) 15% to 95% 10% to 95%
Barometric (mmHg) 427.5 to 805.5 120 to 805.5
Microstream CO
2
module
Item Operating conditions Storage conditions
Temperature (℃) 0 to 40 -20 to 60
Relative humidity (noncondensing) 15% to 95% 10% to 95%
Barometric (mmHg) 430 to 790 430 to 790
Sidestream CO
2
module
Item Operating conditions Storage conditions
Temperature (℃) 5 to 40 -20 to 60
Relative humidity (noncondensing) 15% to 95% 10% to 95%
Barometric (mmHg) 430 to 790 430 to 790
The environmental specifications of unspecified parameters are the same as those of the main unit.