7-9
7.6 Pausing Alarms
If you want to temporarily prevent alarms from sounding, you can pause alarms by pressing the hardkey on the
monitors front. When alarms are paused:
No alarm lamps flash and no alarms are sounded.
No numeric and alarm limit flash.
No alarm messages are shown.
The remaining pause time is displayed in the physiological alarm area.
The alarms paused symbol is displayed in the sound symbol area.
If the time interval of the monitors last shutdown from this starting-up is greater than 2 minutes, the monitor enters into
the alarm paused status as soon as it is turned on. The alarm pause time is fixed to be 2 minutes.
When the alarm pause time expires, the alarm paused status is automatically cancelled and the alarm tone will sound.
You can also cancel the alarm paused status by pressing the hardkey.
You can set the alarm pause time to [1 min], [2 min], [3 min], [5 min], [10 min], [15 min], or [Permanent]. The default
alarm pause time is 2 minutes.
1. Select [Main Menu][Maintenance >>][User Maintenance >>]enter the required password.
2. Select [Alarm Setup >>][Alarm Pause Time] and then select the appropriate setting from the popup list.
7.7 Switching Off All Alarms
If [Alarm Pause Time] is set to [Permanent], the monitor enters into the alarm off status after the hardkey is
pressed. During the alarm off status,
As for physiological alarms, no alarm lamps flash and no alarms are sounded.
As for physiological alarms, no numeric and alarm limit flash.
No physiological alarm messages are shown.
[Alarm Off] is displayed in the physiological alarm area with a red background.
As for technical alarms, no alarms are sounded.
The alarm off symbol is displayed in the sound symbol area.
You can cancel the alarm off status by pressing the hardkey.
WARNING
Pausing or switching off alarms may result in a hazard to the patient. Please be very careful.