12-5
For example, when the clock is enabled, if [Interval] is set to [20min], and then you start NIBP auto measurement at 14:
03, the next measurement will be taken at 14: 20, and the following measurement time will be 14:40, 15:00, and so on.
To enable the clock, in the [NIBP Setup] menu, set [Clock] to [On].
NOTE
The clock function is available only when the auto measurement internal is 5 minutes or more.
12.5.5 Starting a STAT Measurement
1. Select the NIBP parameter window to enter the [NIBP Setup] menu.
2. Select [NIBP STAT]. The STAT mode initiates 5 minutes of continuous, sequential, automatic NIBP measurements.
WARNING
Continuous non-invasive blood pressure measurements may cause purpura, ischemia and neuropathy in
the limb with the cuff. Inspect the application site regularly to ensure skin quality and inspect the extremity
of the cuffed limb for normal color, warmth and sensitivity. If any abnormity occurs, move the cuff to
another site or stop the blood pressure measurements immediately.
12.6 Understanding the NIBP Numerics
The NIBP display shows numerics only as below. Your display may be configured to look slightly different.
1. Time of last measurement
2. Time remaining to next measurement
3. Measurement mode
4. Unit of pressure: mmHg or kPa
5. Systolic pressure
6. Diastolic pressure
7. Mean pressure obtained after the measurement and cuff pressure obtained during the measurement
3
4
5
6
7
1
2