11-5 
 
Sat-Seconds= Points × Seconds 
 
Only when the Sat-Seconds limit is reached, the monitor gives a Sat-Seconds alarm. For example, the figure below 
demonstrates the alarm response time with a Sat-Seconds limit set at 50 and a low SpO
2
 limit set at 90%. In this example, 
the patient % SpO
2 
drops to 88% (2 points) and remains there for 2 seconds. Then it drops to 86% (4 points) for 3 seconds, 
and then to 84% (6 points) for 6 seconds. The resulting Sat-Seconds are: 
% SpO
2
  Seconds  Sat-Seconds 
2×  2=  4 
4×  3=  12 
6×  6=  36 
Total Sat-Seconds=  52 
 
After approximately 10.9 seconds, a Sat-Second alarm would sound, because the limit of 50 Sat-Seconds would have 
been exceeded. 
 
 
 
Saturation levels may fluctuate rather than remaining steady for a period of several seconds. Often, the patient % SpO
2
 
may fluctuate above and below an alarm limit, re-entering the non-alarm range several times. During such fluctuation, 
the monitor integrates the number of % SpO
2
 points, both positive and negative, until either the Sat-Seconds limit is 
reached, or the patient % SpO
2
 re-enters the non-alarm range and remains there.   
 
 
11.5.7 Changing the Speed of the Pleth Wave 
In the [SpO
2
 Setup] menu, select [Sweep] and then select the appropriate setting. The faster the waveform sweeps, the 
wider the waveform is. 
 
 
11.5.8 Setting the Alarm Level for SpO
2
 Sensor Off Alarm 
Select [Alarm Setup >>] from the [User Maintenance] menu. You can set the [SpO2SensorOff Lev.] in the popup 
menu. 
 
 
Seconds 
SpO
2