A-13
awRR High (low limit + 2) to 150 rpm
1 rpm
awRR Low 0 to (high limit - 2) rpm
Microstream CO
2
Module
Standard Meet standard of ISO 21647
CO
2
Measurement range 0 to 99 mmHg
Accuracy*
0 to 38 mmHg:
39 to 99 mmHg:
±2 mmHg
±5% of the reading+0.08% of (the reading-38)
Accuracy drift Meet the requirement for measurement accuracy within 6 hours
* Accuracy applies for respiration rate up to 80 rpm. For respiration rate above 80 rpm, the accuracy is 4 mmHg or ±12% of
the reading, whichever is greater. for EtCO
2
exceeding 18 mmHg. In the presence of interfering gases, the above accuracy is
maintained to within 4%.
Resolution 1 mmHg
Sample flow rate
7.5
15
50
+
ml/min
Initialization time 30 s (typical)
Response time
Measured with a FilterLine of standard length:
2.9 s (typical)
4.5 s (Maximum)
(The response time is the sum of the rise time and the delay time)
Rise time: <190 ms (10% to 90%)
Delay time: 2.7 s (typical)
awRR measurement range 0 to 150 rpm
awRR measurement accuracy
0 to 70 rpm:
71 to 120 rpm:
121 to 150 rpm:
±1 rpm
±2 rpm
±3 rpm
Apnea alarm time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Alarm limit Range Step
EtCO
2
High (low limit + 2) to 99 mmHg
1 mmHg
EtCO
2
Low 1 to (high limit - 2)mmHg
FiCO
2
High 1 to 99 mmHg
awRR High (low limit + 2) to 150 rpm
1 rpm
awRR Low 0 to (high limit - 2) rpm