3-13
IBP Pressure Calibration
Tools required:
Standard sphygmomanometer
Balloon pump
Tubing
T-shape connector
1. Connect the equipment as shown below
2. Zero the transducer. After a successful zero, open the stopcock to the manometer.
3. In the main menu, select [Maintenance >>][User Maintenance >>]enter the
required password. In the [Maintain IBP] menu, enter calibration values for calibrated
pressures.
4. Inflate using the balloon pump until the mercury in the manometer reaches to the
preset calibration pressure value.
5. Adjust the preset calibration value until it equals to the reading on the manometer.
6. Select the [Calibrate] button to the left of the calibrated pressure. The monitor starts a
calibration.
7. When the calibration is completed, the message [Calibration Completed!] is
displayed. If the calibration failed, a prompt message will be displayed.
Manometer
T-shape connector
3-way stopcoc
k
Pressure transducer
Pressure adapter cable
Monitor