4-9
4.6.9 Network Related Problems
Symptoms Possible Cause Corrective Action
Incorrect LAN cable
connection
Check LAN cable connection. LAN cable
shall not be longer than 50 m.
Frequent dropouts and
network disconnects
Incorrect IP address
configuration
Check for IP address conflict. Reconfigure
IP address.
Incorrect LAN cable
connection
Check LAN cable connection. LAN cable
shall not be longer than 50m.
Excessive requests for
viewing the patient
monitor at the same
time
A patient monitor can only be viewed by 4
other patient monitors at the same time. The
excessive view requests system will be
ignored.
The patient monitor is
connected to a LAN
but cannot view other
patients in the [View
Other Patient]
window.
Incorrect IP
configuration
Check for IP address conflict. Reconfigure
IP address.