3.4.10 AG Tests
AG Performance Test
Tool required:
Gas bottle with a certain standard anesthetic gas. Gas concentration should meet
AA>1.5%, of which AA represents an anesthetic agent
T-shape connector
Tubing
1. Enter [Gas Setup] menu and set [Operating Mode] to [Measure].
2. After the AG module warmup is ready, block the gas inlet of the AG module. Alarm
message [AC Airway Occluded] is displayed.
3. Connect the equipment as shown below.
4. Open and close the gas valve repeatedly at an interval of 3 seconds. The AA value
displayed should meet product specification.
Gas valve
Gas bottle
Tubing
Open to the air
Monitor