5-21
IBP/C.O. communication cable
5.6.2 Removing the IBP/C.O. Module
Unplug the communication cable to the multi-parameter board. Then unscrew the
M3×6 screw to take out the IBP/C.O. module.
NOTE
z When reinstalling, place the board according to the positioning nubs.
z When reinstalling, do not mix up the IBP3 and IBP4 connectors. Place the cables in
the way as shown below. Do not place them above the plastic pillar.
IBP4
The positioning
nubs
Plastic
pillar
Screw