3-22
3.8 Touchscreen Calibration
1. In the main menu, select [Maintenance >>]→[User Maintenance >>]→enter the
required password and then select [Cal. Touchscreen] from the popup menu.
2.
will, in turn, appear at different positions of the screen.
3. Select each
as it appears on the screen.
4. After the calibration is completed, the message [Screen Calibration Completed!] is
displayed. Select [Ok] to confirm the completion of the calibration.
3.9 Recorder Check
1. Print ECG waveforms. The recorder should print correctly and printout should be
clear.
2. Set the recorder to some problems such as out of paper, paper jam, etc. the patient
monitor should give corresponding prompt messages. After the problem is removed,
the recorder should be able to work correctly.
3. Switch automatic alarm recording for each parameter ON and then set each
parameter’s limit outside set alarm limits. Corresponding alarm recordings should be
triggered when parameter alarms occur.