3.11 Program Upgrade
You can upgrade the software of the patient monitor by downloading the upgrade software
through network. At the same time, you can view the upgrade status log. The upgrade
software can directly run on a portable PC or desktop PC. Through the network or by
connecting the patient monitor to a PC via a crossover network cable, you can upgrade the
following programs:
Boot program
System program
Multilingual library
BMP files (including screen icons, start-up screens, standby screens)
General configurations (including passwords, company logo)
System functional configuration
FPGA program
Module programs: M54A module, M02B module, M03B module, Power module
For details, refer to help and instructions for program upgrade.
CAUTION
z Disconnect the patient monitor from the patient and make sure the important data
are saved before upgrade.
z Do not shut down or power off the equipment when upgrading the boot program.
Otherwise, it may cause the equipment to break down.
z Program upgrade should be performed by qualified service personnel only.
NOTE
z After upgrading the boot program, re-upgrade the system program and other
programs to ensure compatibility.
z Make sure the version of the upgrade package is you desired one. If you want to
obtain the latest upgrade package, contact Mindray Customer Service Department.