3-14
3.4.7 SpO
2
Test
Tool Required: Patient simulator.
1. Connect the patient simulator to the SpO
2
connector on the monitor.
2. Select the model and the manufacturer of the SpO
2
module under test, and then
configure the patient simulator as follows: SpO
2
96%; PR 80 bpm.
3. The displayed SpO
2
and PR values should be within the ranges listed below.
SpO
2
PR (bpm)
Mindray 96% ± 2% 80 ± 3
Masimo 96% ± 2% 80 ± 3
MAX-A, MAX-AL, MAX-N, MAX-P,
MAX-I, MAX-FAST
96% ± 2%
OxiCliq A, OxiCliq N, OxiCliq P, OxiCliq I 96% ± 2.5%
D-YS, DS-100A, OXI-A/N, OXI-P/I 96% ± 3%
Nellcor
MAX-R, D-YSE, D-YSPD 96% ± 3.5%
80 ± 3
3.4.8 C.O. Test
Tool required: Patient simulator
1. Connect the patient simulator to the C.O. connector on the monitor.
2. Set the patient simulator to output blood temperature (BT) 37ºC. The displayed
temperature value should be 37 ± 0.1ºC.
3.4.9 Temp Test
Tool required: Resistance box
1. Connect the two pins of any Temp connector on the monitor to the two ends of the
resistance box using 2 wires.
2. Set the resistance box to 1354.9Ω (corresponding temperature is 37ºC). The displayed
value should be 37 ± 0.2ºC.