5-26
NOTE
z When reinstalling the pump, aim the notch of the pump toward the positioning nub
on the support rack.
z When reinstalling the valves, do not mix up the installing positions of the
slow-release valve and fast-release valve.
z When reinstalling, connect all the tubings in position.
5.6.6 Disassembling the Main Support Assembly
1. Unplug the power input cable (from power board to power management and interface
board) and power adapter board cable (from battery board to power management and
interface board).
NOTE
z Release the clamps before unplugging the cables.
Notch
Positioning nub