5-23
5.6.4 Removing the Multi-parameter Board
1. Release the two clamps. Unplug the fast-release valve cable, slow-release valve cable
and pump cable.
2. Unplug the tubings.
3. Unscrew the three M3×6 screws and pull up the multi-parameter board.