5-24
4. Unscrew the two M3×6 screws to take out the ECG shield.
NOTE
z Release the clamps before unplugging the cables.
z Make sure all the connectors have been removed before taking out the
multi-parameter board. Take out the board carefully to avoid damage.
z When reinstalling the ECG shield, place it according to the positioning slot to avoid
the short circuit between the shield and other components.
z Set the jumper as follows. S5 is OFF for CO2 and ON for AG.