3.4.6 IBP Tests
IBP Performance Test
Tool required:
Patient simulator
1. Connect the patient simulator to the pressure connector on the monitor.
2. Set the pressure value of patient simulator to 0.
3. In the setup menu for the pressure (e.g. Art), select [Art Zero >>]→[Zero] to start a
zero calibration.
4. Configure the patient simulator as P(static)=200 mmHg.
5. The displayed value should be 200 ± 2 mmHg.
6. If the value is beyond tolerance, perform the pressure calibration.