5-19
To remove the NIBP connector:
1. Unscrew the connector with a sharp nose plier. Then carefully take out the outer part.
Do not lose the steel balls.
2. Take out the plastic cover from the outer part with a tweezer.
NOTE
z When reinstalling, insert a magnetic phillips screwdriver into the inner part.
Place the four steel balls into the holes and then install the plastic cover. All the
steel balls must be placed.
z When reinstalling, aim the groove of the inner part toward the salient of the front
panel.