4 Setup
Operator’s Manual 4 - 47
If you have any questions, please contact the service engineer or your sales representative.
Setup Export Log Export the log.
Self Test Perform system self-test and restart the machine.
Recover To recover the system.
Enter Windows For password, please contact your service engineer.
Prepay Installment Display the prepay installment information.
Start/Stop WinProxy Cursor debug interface.
Touch Screen
Calibration
To calibrate the touch screen.
Needles List Replace Update the list of needles that support eSpacial Navi function.
NOTE:
The list of needles must be release version provided by
Mindray.
Needles List Recover Restore the list of needles that support eSpacial Navi function
to the last version.
Needles List Factory Restore the list of needles that support eSpacial Navi function
to the factory version.
Enable Tracking Enable the transducer and patient tracking function:
• Find transducers used with a patient by patient ID.
• Find patients who used a transducer by transducer SN.
After enabling the tracking function, you can set the maximum
number of days the data can be retained.
Fast Startup To speed up the startup process.
NOTE:
If McAfee has been installed, please disable this function
to avoid McAfee failure.
Item Description