EasyManuals Logo

Mindray MX7 User Manual

Mindray MX7
338 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #146 background imageLoading...
Page #146 background image
6 - 30 Operator’s Manual
6 Image Acquisition
6.15.3 Select Image and Cardiac Cycle
You can select images with a better image quality so as to guarantee the analysis result.
Switch the cine file
1. Tap [Review] to enter the review state on touch screen.
2. Double-click the target file.
3. The system closes current displayed file and switch to the newly selected file.
Switch cardiac cycle within the cine file
Tap [Circle] to select when opening a cine which includes multiple cardiac cycles.
6.15.4 Myocardial Boundary Tracing
Tracing
The system provides 2 kinds of tracing method for 2 kind of sections. Long axis section (A4C,
A2C, ALAX): 3-point method and manual tracing method are both available. Short axis section
(PSAX B, PSAX M, PSAX AP): only manual tracing is available.
• 3-point method
As shown in the following figure, after operation by pressing <Set> to place 3 points on the
image, the system generates the trace automatically.
• Manual trace method
Press <Set> and move the cursor by using the trackball/trackpad along the boundary to add the
trace points gradually, after trace is finished, press <Set> twice to finish tracing.
NOTE:
At least 6 points should be determined by you before the system generates automated trace.
Press <Set> to make the traces on the image clockwise or anticlockwise.
Retracing
If current trace is not satisfactory, tap [Reload] on the touch screen to clear the trace and to start
another tracing.
During the tracing drawing, press <Clear> to clear already traced drawing.
Make fine adjustment to the trace
You can make fine adjustments to the trace after it is completed.
Perform the following procedure:
1. Under tracing curve adjusting status, the cursor turns into .
2. Move the cursor to the editable point, press <Set>.
1
2
3

Table of Contents

Other manuals for Mindray MX7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray MX7 and is the answer not in the manual?

Mindray MX7 Specifications

General IconGeneral
Display Size15.6 inches
Transducer Ports3
PortabilityPortable
TypeUltrasound System
Display TypeLCD
Battery Life120 min
Imaging ModesB, M, Color Doppler, Power Doppler, Pulsed Wave Doppler
ConnectivityWi-Fi, Ethernet, USB
ApplicationsVascular, Cardiology, OB/GYN, Urology
Probe Connectors3
Operating SystemWindows