4 - 36 Operator’s Manual
4 Setup
Worklist Setting
DICOM service setting for Worklist is described as follows:
MPPS Preset
MPPS setting items are described as follows:
Strategy Name Set the DICOM strategy.
Item Description
Item Description
Device Name After you set the server (s) in DICOM Server Setting screen, the name
(s) will appear in the drop-down list, select the name of the Worklist
server.
Service Name Default is server-Worklist, and it can be modified.
AE Title Application Entity title. It is consistent with that of the Worklist server.
Port DICOM communication port, 104 by default. The port should be
consistent with that of the Worklist server port.
Maximum Retries Reserved feature.
Interval Time(s) Reserved feature.
Timeout Refers to time after which the system will stop trying to establish a
connection to the service.
TLS Transport Layer Security. Select whether to encrypt the data during
network transportation.
Strategy Name Set the DICOM strategy.
Remove Attributes(0) Set what DICOM element(s) that will not be used in worklist query.
Item Description
Device Name After you set the server (s) in DICOM Server Setting, the name (s) will
appear in the drop-down list, select the name of the MPPS server.
Service Name Default is server-MPPS, and it can be modified.
AE Title Application Entity title. It should be consistent with that of the MPPS
server.
Port DICOM communication port, 104 by default. The port should be
consistent with that of the MPPS server.
Maximum Retries It starts retrying if it fails to send DICOM task to the server. The retry
entry times should be this value.
Interval Time(s) Reserved feature.
Timeout Refers to the amount of time after which the system will stop trying to
establish a connection to the service.
TLS Transport Layer Security. Select whether to encrypt the data during
network transportation.

Table of Contents

Other manuals for Mindray MX7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray MX7 and is the answer not in the manual?

Mindray MX7 Specifications

General IconGeneral
Display Size15.6 inches
Transducer Ports3
PortabilityPortable
TypeUltrasound System
Display TypeLCD
Battery Life120 min
Imaging ModesB, M, Color Doppler, Power Doppler, Pulsed Wave Doppler
ConnectivityWi-Fi, Ethernet, USB
ApplicationsVascular, Cardiology, OB/GYN, Urology
Probe Connectors3
Operating SystemWindows

Related product manuals