EasyManuals Logo

Mindray SV300 User Manual

Mindray SV300
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #155 background imageLoading...
Page #155 background image
11-3
Parts
Recommended
frequency
Cleaning Disinfection
A B C D
Ventilator Housing
Ventilator external surface (including
housing, power cord, supply gas hose)
Each patient A or D
Trolley and support arm Each patient A or D
Touch screen Each patient A or D
Fan dust filter
Every four weeks/as
necessary*
B
Main unit air outlet dust filter
Every four weeks/as
necessary*
B
Air intake dust filter
Every four weeks/as
necessary*
B
Ventilator inspiration safety valve assembly
Inspiration safety valve assembly as necessary* B or C
Ventilator expiration valve assembly
Expiration valve membrane (silicone) Each patient/weekly B or C
Expiration valve assembly (except
membrane)
Each patient/weekly B or C
Ventilator patient tubing (reusable)
Patient tubing (including water
trap, Y piece, adapter)
Each patient/weekly B or C
Other
Mainstream CO
2
sensor Each patient/weekly
Refer to the cleaning and
disinfection methods provided by
the mainstream CO
2
vendor.
SpO
2
sensor Each patient/weekly Refer to the cleaning and
disinfection methods provided by
the attached package insert.
SpO
2
sensor cable Each patient/weekly Refer to the cleaning and
disinfection methods provided by
the attached package insert.
Nebulizer Each patient/weekly
Refer to the cleaning and
disinfection methods provided by
the nebulizer vendor.
Humidifier Each patient/weekly
Refer to the cleaning and
disinfection methods provided by
the humidifier vendor.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SV300 and is the answer not in the manual?

Mindray SV300 Specifications

General IconGeneral
Waveform DisplayPressure, Flow, Volume
Battery BackupUp to 2 hours
Ventilation ModesVCV, PCV, SIMV
Tidal Volume Range20 to 2000 mL
FiO221 to 100%
Inspiratory Time0.2-5.0 s
Monitoring ParametersPaw, Ppeak, Pmean, Pplat, PEEP, VTI, VTE, MV, f, FiO2, I:E
AlarmsHigh/Low Airway Pressure, High/Low Minute Volume, High/Low Tidal Volume, High/Low Respiratory Rate, High/Low FiO2, Apnea, Power Failure
Power SupplyAC 100-240V, 50/60Hz
Peak Flow180 L/min
Trigger Sensitivity (Flow)0.5-10 L/min