11-4
Parts
Recommended
frequency
Cleaning Disinfection
A B C D
Cleaning methodsWipe and Bath Immersion:
Wipe: wipe with a damp cloth immersed in alkalescent detergent (soap water, etc.) or alcohol
solution and then wipe off the remaining detergent with a dry lint free cloth.
Immersion: flush with water first and then immerse it in alkalescent detergent (soap water, etc.)
(water temperature 40ºC recommended) for approximately three minutes. Finally clean with water
and dry completely.
Disinfection methods:
A: Wipe: wipe with a damp cloth immersed in medium- or high-efficiency detergent and then wipe
off the remaining detergent with a dry lint free cloth.
B: Immersion: immerse it in medium- or high-efficiency detergent for more than 30 minutes
(recommended time). Then clean with water and dry completely.
C: Steam autoclave at 134ºC for 10 to 20 minutes (recommended time).
D: Ultraviolet radiation for 30 to 60 minutes (recommended time).
As necessary*: shorten the cleaning and disinfection intervals if the equipment is used in
dusty environment to ensure that the equipment surface is not covered by dust. Clean and
disinfect the inspiration safety valve assembly only when the patient’s exhaled gas may
contaminate the inspiratory limb. For disassembling and installation methods, refer to 10.2.2.
The table below lists the cleaning and disinfecting agents and autoclaving process that may
be used on the ventilator.
Name Type
Ethanol (75%) Moderately efficient disinfectant
Isopropanol (70%) Moderately efficient disinfectant
Glutaraldehyde (2%) Highly efficient disinfectant
Ortho-Phthalaldehyde disinfectant (such as
Cidex
®
OPA
Highly efficient disinfectant
Soap water (pH value of 7.0~10.5) Rinsing agent
Clean water Rinsing agent
Steam autoclave* Highly efficient disinfection
Steam autoclave*: The recommended temperature of this disinfection method is 134ºC
(273ºF).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SV300 and is the answer not in the manual?

Mindray SV300 Specifications

General IconGeneral
Waveform DisplayPressure, Flow, Volume
Battery BackupUp to 2 hours
Ventilation ModesVCV, PCV, SIMV
Tidal Volume Range20 to 2000 mL
FiO221 to 100%
Inspiratory Time0.2-5.0 s
Monitoring ParametersPaw, Ppeak, Pmean, Pplat, PEEP, VTI, VTE, MV, f, FiO2, I:E
AlarmsHigh/Low Airway Pressure, High/Low Minute Volume, High/Low Tidal Volume, High/Low Respiratory Rate, High/Low FiO2, Apnea, Power Failure
Power SupplyAC 100-240V, 50/60Hz
Peak Flow180 L/min
Trigger Sensitivity (Flow)0.5-10 L/min