11-18
To disassemble the humidifier from the ventilator:
1. Disconnect the tubes from the humidifier.
2. Remove the screw.
3. Lift up the humidifier to remove it from the humidifier bracket fixed seat.
To install the humidifier onto the ventilator:
1. Align the humidifier mounting plate and the slot, and slide the humidifier in.
2. Tighten the screw.
3. Mount the filters onto the inspiratory and expiratory ports.
4. Connect the inspiratory filter to the humidifier inlet via the tube.
5. Connect the humidifier outlet to the water trap via the tube. Then, connect the water trap
to the Y piece via the tube.
6. Connect the expiratory filter to the water trap via the tube. Then, connect the water trap
to the Y piece via the tube.
7. Place the patient tubing onto the support arm hook.