MP 26 Transmitter Installation and Calibration Manual
Minebea Intec
EN-29
Step
Display Line 2
Operation
Description
Adjusting the Dead Load Calibration
5.
kg
Press for more than 3 sec
to display InSt. Press to
go to SEt. Use or and
move to CAL
After finishing the full-scale
calibration, the dead weight is
shown in display1.
Press to move into
calibration mode. Then, the
display shows InL.1
Here, calibration can be
adjusted.
In this menu level, the
display alternates between
OFF and InL.1.
Now set the dead load. Use
or to activate the
measured value. Note down
the measured value as CtL.
Press .
Now the display moves to
InH.1. Press until you
reach
SCALinH.1
Set the dead load by removing
any weight apart from the dead
load from the system
XXXX only shows the measured
value in percent. This finishes
setting the dead load.
SCALinH.1
Set the calibration weight
using or .
Then, press until End is
reached.
Set the adjusted calibration
value as CH = (PreL - CtL)+ PreH.
Afterwards, the dead load is
adjusted to the current value.