MP 26 Transmitter Installation and Calibration Manual
Minebea Intec
4.3.3. Dead Load Correction
The dead load can be adjusted without the need of an entire calibration. The steps are
explained as follows.
kg
Press more than 3 sec it
will display InSt. press to
go to SEt. use or and
move to SCAL.
If the initial settings have been
configured earlier then we can
directly go to Weighing mode.
SCAL
Press to move into
calibration by mode.
Now the display shows InL.1
Here calibration by value can be
performed.
Note down the previous Calibration values
InL.1
Note down the InL.1 value
consider this as PreL.
Press twice to go to
InH.1
In this menu level, the display
alternates between InL.1 and the
set value. The value displayed
here shows the previous
calibration’s dead load. Note the
value.
InH.1
Note down the InH.1 value
consider this as PreH.
Press until the display
shows dOnE and moves to
End. Press to go to
normal display mode.
In this menu level, the display
alternates between InH.1 and set
value. The value displayed here
shows the previous calibration’s
max value.