MP 26 Transmitter Installation and Calibration Manual
Minebea Intec
EN-34
4.5. Tare Setting
The tare function is used to tare the weight on the scale. The scale displays zero. Then, the net
weight can be measured (net = gross – tare). The gross weight can be found by switching off
the tare function.
The procedure for setting the tare is:
1. Press the for more than 3 sec. It will lead to Inst.
2. Use or to move to COnF and press key.
3. It will display FUnC. Press the key.
4. Press key to move to FnC.2 and use or to set it to 1.
Function
Value
Description
FnC.2
Function 1
0
No function
1
Tare setting
1. Press to move to done and InP.1.
2. Use or to move to LOGI and press .
3. Press until tArA is reached.
4. Use or and set the logic required for tarring the scale and confirm with the
key.
tArA
Tare function (function must be activated
(CONF /FUNC / Fnc.2= 3))
0
Interface only
2
di1 switches
7
Limit 1 switches
8
Limit 2 switches
9
Limit 3 switches
10
Enter/Inc keys switch
11
Enter/Dec keys switch
5. dOnE is displayed. Move to rESL. Press to move to Othr. Use or to move to
End and confirm with .