MP 26 Transmitter Installation and Calibration Manual
Minebea Intec
EN-48
7.5. Engineering Front Interface
Connection to the controller front via a PC adapter (see ‘Additional Accessories’). The
Engineering software enables the MP 26 to be
configured, parameters set, and operated.
7.6. Environmental Conditions
7.6.1. Protection mode
Front panel:
IP 20
Housing:
IP 20
Terminals:
IP 20
7.6.2. Permissible temperatures
For specified accuracy:
-1055°C
Warm-up time:
<
20 minutes
Temperature effect:
<
0.08 %/10 K
Operating limits:
-2060°C
Storage:
-3070°C
7.6.3. Humidity
Max. 95 %, 75 % yearly average, no condensation
7.6.4. Shock and vibration
Vibration test Fc (DIN EN 60 068-2-6)
Frequency:
10…150 Hz
Unit in operation:
1 g or 0.075 mm
Unit not in operation:
2 g or 0.15 mm
Shock test Ea (DIN EN 60 068-2-27)
Shock:
15 g
Duration:
11 ms
7.6.5. Electromagnetic compatibility
Meets the test requirements for instruments in industrial areas.
Interference radiation
Within the limits for class A instruments.
Immunity to interference
Meets EN 61326-1 for continuous, unattended operation.