MP 26 Transmitter Installation and Calibration Manual
Minebea Intec
4.3.4. Calibration by Value
Calibration by value can be done in an indirect way. Switch to SCAL mode if the initial
settings have been configured previously or configure the initial settings as explained above
and then switch to SCAL mode.
An example for calibration by value is given in Chapter 4.6. Please refer to it for detailed
information.
Calibration by value is done indirectly. It is a theoretical calibration. The calibration is done
based on the assumption that the dead load is zero. From this, knowing the dead load weight,
adjust the zero point to the dead load. The calibration point is shifted accordingly to the zero
point shifts, thereby calibrating the system to its dead load weight.
Press for more than 3 sec
to display InSt. Press to
go to SEt. Use or and
move to SCAL.
If the initial settings have been
configured earlier, then, go
directly to SCAL mode.
Press to move to
calibration by value.
Now, the display shows
Here, calibration by value can be
performed.
Calibration for standard weight
Set the dead load to zero
using the or arrow key.
Then press .
Now the display shows
OuL.1.
In this menu level, the display
alternates between InL.1 and set
value. Enter zero percent as dead
load.