MP 26 Transmitter Installation and Calibration Manual
Minebea Intec
6.2. Error List
Internal error, cannot be
corrected
EEPROM
Return device to
manufacturer
Internal error, resettable
supply cables separate.
Protect contactors by
means of RC snubber
resettable
configuration
of configurations and
parameters
Code number and
hardware not identical
Lim.2
alarm
1/2/3 exceeded
Note:
*) “FAIL” is shown on the process value display.
Latched alarms Lim1/2/3 (E-element displayed) can be acknowledged, i.e. reset via digital alarm di1.
If an alarm is still pending, i.e. unless the error cause was removed (E display blinks), latched alarms cannot be
acknowledged and reset.
Change to error status 1 after error removal
Change to error status 0 after acknowledgement in error list 0
Not visible, except during acknowledgement
If sensor errors are not on the error list any more after error correction without manual reset in the error list,
suppression via MP26/10 is possible by setting ILat.
CONF othr ILat 1 (blocked)
This setting has no effect on limit values Lim.1…3 configured for storage.
Detection and display of sensor and wiring errors