Cause:   The pump has detected excessive positive pressure between the pump and the 
  IV Source.
Resolution:   Press PAUSE to clear alarm screen. Check for excessive pressure on the IV bag 
  and relieve it. When ready to begin infusion, select “RESUME.”
LCD Display:  ***ALARM*** REPLACE SET 3
Alarm:   Replace Set
Cause:   The pump has detected that the disposable administration set must be 
 replaced.
Resolution:   Replace the disposable administration set, prime the set if necessary, and when 
  ready to begin infusing, select “RESUME” to continue the infusion.
LCD Display:  ***ALARM*** REPLACE SET 4
Alarm:   Replace Set
Cause:   The pump has detected excessive pressure in the disposable administration set.
Resolution:   Make sure that all the IV clamps are open. Relieve pressure, open and close 
  pump door.
LCD Display:   ***ALARM*** SET NOT INSTALLED
Alarm:   Administration Set Not Installed.
Cause:   The administration set has not been installed or has been installed improperly.
Resolution:   Install appropriate Moog Curlin Infusion Administration set for desired 
  therapy. Use ONLY Moog Curlin Infusion Administration sets. If the 
  blue tubing guide on the administration set has been damaged in any way, 
  replace the entire administration set. See Chapter 2, Figures 2.9 and 2.10 for 
  proper installation of administration set.
LCD Display:   ***ALARM*** UNATTENDED PUMP
Alarm:   No key presses have occurred while the pump was paused
Cause:   When a therapy is in progress and a pump is placed in a paused state and 
  left unattended for more than two minutes, this alarm will sound. If the 
  prime menus are left unattended for one minute, this alarm will sound
Resolution:   Press the PAUSE key to clear the alarm screen. If required, return to the 
  screen displayed at the time of the alarm and complete the activity that was in
  progress, or resume the infusion, or turn the pump off. 
LCD Display:   ***ALARM*** UP OCCLUSION 
Alarm:   Negative pressure detected between the pump and IV reservoir.
Cause:   Pump has detected excessive negative pressure in the administration set 
  between the medication reservoir or IV bag (“up stream”) and the pump.
Troubleshooting