Resolution: Pause the pump. Check administration set from IV bag to pump for
potential causes of occlusion. Open the door of the pump and remove the
administration set from the pumping chamber. If an area of the tubing is
flattened, massage that area to a more rounded shape. If a clamp is in place,
open it; if tubing is kinked, straighten it. Replace administration set back
in the pump. When ready to begin infusing, select “RESUME.” If the alarm
continues, use a new administration set or if the IV bag is almost empty,
replace it.
LCD Display: COMM TIMEOUT
Alarm: Communication Timeout occurred during Pump/CMS communication.
Cause: The communication link between the pump and the PC running CMS was
broken (i.e., the cable between them was unplugged) during a CMS upload to
the pump
Resolution: Check the cable connection between the pump and the PDA or PC.
LCD Display: INFUSION COMPLETE
Alarm: Infusion Complete
Cause: Programmed “Amount to be Infused” is depleted to 0.
Resolution: If the therapy is to be repeated, press PAUSE, add another IV bag or
reservoir and select the “REPEAT Rx” function.
If the therapy will not be repeated, press PAUSE key and turn the pump off
by pressing the OFF key.
LCD Display: STUCK BOLUS KEY
Alarm: Bolus Key is stuck.
Cause: The Remote Bolus Cord is not functioning correctly
Resolution: Replace the Remote Bolus Cord.
LCD Display: ***ALARM *** LOW MOTOR SUPPLY
Alarm: Low Motor Supply Voltage
Cause: The voltage to the motor supply is too low
Resolution: Connect the AC adaptor or replace the batteries.
LCD Display: ***ALARM *** MOTOR STALLED
Alarm: Motor Stalled
Cause: The motor is not turning at the expected rate.
Resolution: Connect the AC adaptor or replace the batteries and resume the infusion.
LCD Display: ***ALARM *** INVALID UNITS FOR PUMP CONFIGURATION
Alarm: Invalid Units for Pump Configuration.