Resolution: Remove the remote bolus cord from the pump if connected. Refrain from
pressing the BOLUS key.
LCD Display: ***ALERT *** NUMBER OF DOSES INVALID
Alert: Number of doses entered exceeds maximum for therapy
Cause: The number of doses per bag in Intermittent therapy exceeds the allowable
maximum.
Resolution: Recheck data entry. The number of doses requested must be less than or equal
to the maximum for the therapy.
LCD Display: ***ALERT *** PERFORM ROUTINE MAINTENANCE
Alert: Maintenance Date (displayed only if date has arrived when pump is turned
on)
Cause: Current Date is beyond the maintenance date.
Resolution: Return pump for routine maintenance.
LCD Display: ***ALERT *** RAMP INVALID FOR AMT/RATE
Alert: Ramp time entered is invalid for amt/rate already programmed
Cause: Ramp value entered for TPN is unacceptable given the desired rate and
amount programmed.
Resolution: Recheck all data entered for prescription. If programming a new TPN therapy
(“NEW PROGRAM” selected) and the ramp time is a requirement, enter the
ramp time(s) and total time first followed by the next governing parameter
(rate or volume to be infused).
LCD Display: ***ALERT *** RATE EXCEEDS DELTA RATE/DELTA TIME
Alert: Rate entered exceeds the delta time limits
Cause: The rate entered falls outside of the allowable range for the delta rate / delta
time limits programmed.
Resolution: Recheck data entry or wait more time elapses then attempt to change the rate.
Rate selected cannot exceed the range placed on medLMTS
.
LCD Display: ***ALERT *** RUNNING, PAUSE PUMP PRIOR TO PRESSING OFF
Alert: Pause Infusion before turning off
Cause: The OFF key was pressed while the pump was infusing or a running screen
displayed.
Resolution: Press PAUSE key before pressing OFF key to stop an infusion.
LCD Display: ***ALERT *** Rx IN PROGRESS, YES to Repeat – NO to Return
Alert: “REPEAT Rx” selected while a therapy is already in progress
Cause: “REPEAT Rx” was selected and there is a therapy currently in progress.