Chapter 3
Specifications
Table 3-1. Performance Specifications
GAS TYPES
Combustibles, Oxygen & Toxics
TEMPERATURE
RANGE
TOXICS &
OXYGEN
OPERATING
RANGE:
0 to 40
o
C (32 to +104
o
F)
*EXTENDED
RANGE:
-20 to +50
o
C (-4 to +122
o
F)
HYDROGEN
OPERATING
RANGE:
10 to 40
o
C (50 to 104
o
F)
*EXTENDED
RANGE:
-20 to +50
o
C (-4 to +122
o
F)
AMMONIA
OPERATING
RANGE:
10 to 40
o
C (50 to 104
o
F)
*EXTENDED
RANGE:
-10 to +50
o
C (-14 to +122
o
F)
CATALYTIC
COMBUSTIBLES
SINGLE
MODULE:
-40 to +50
o
C (-40 to +122
o
F)
DUAL MODULE:
-40 to +90
o
C (-40 to +194
o
F)
IR
COMBUSTIBLES
-20 to +50
o
C (-4 to +122
o
F)
IR CARBON
DIOXIDE
OPERATING
RANGE:
0 to 40
o
C (32 to +104
o
F)
*EXTENDED
RANGE:
-10 to +50
o
C (-14 to +122
o
F)
DRIFT
ZERO DRIFT
Less than 5%/year, typically
SPAN DRIFT
Less than 10%/year, typically
NOISE
Less than 1% FS
*EXTENDED RANGE = The sensor may not meet all of the accuracy parameters listed.
Calibrate within operating range.
Chapter 3, Specifications