EasyManua.ls Logo

MSA Ultima - Chapter 4 Maintenance; General; Replacing the Sensor

MSA Ultima
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4
Maintenance
General
The only routine maintenance item is the sensing element itself,
which has a limited lifetime. When the Ultima Gas Monitor’s
read-out indicates that the sensor must be changed, there is very
little sensor lifetime remaining. It is good practice to obtain a
replacement sensing element before the sensing element within
your unit becomes inoperative.
The Ultima Gas Monitor is constantly performing a self-check.
When a problem is found, it displays the appropriate error code.
(TABLE 4-1, "Troubleshooting Guidelines"). When any of these
errors are detected within the unit, the output signal falls to its
lowest possible output.
For frequency output units: lowest possible output is 0 KHz
For 4 to 20 milliamp output models: lowest output is <3.75 mA
The "Change Sensor" indication is not an error and does not
affect the output.
Replacing the Sensor
Typically, the Ultima Monitor LCD display shows a maintenance
message when the sensor is due for replacement (FIGURE 4-1).
Hydrogen Cyanide and Hydrogen Chloride Sensors
NOTE: This NOTE and the following CAUTION apply to the
Ultima Series sensors for Hydrogen Cyanide (HCN)
and Hydrogen Chloride (HCL) only.
Figure 4-1.
Change Sensor Flag
Chapter 4, Maintenance
4-1

Table of Contents

Other manuals for MSA Ultima

Related product manuals