43 EDH0170FE1010 – 09/98
LDS1000 Controller for Laser Diode Sensors
10.5.04 MS Read Acquisition Status
Use h MNL j RMT j ACT
Syntax MS
Parameter None.
Description Read acquisition status. The return value is an ASCII character
the code of which should be translated into binary.
Bit 0= 0: Correct power received.
= 1: Incorrect power received.
Bit 1= 0: Correct position.
= 1: Incorrect position.
Bit 2= 0: Acquisition not started up.
= 1: Acquisition started up - TRIGG wait.
Bit 3= 0: No acquisition in progress.
= 1: Acquisition in progress.
Bit 4= 0
Bit 5= 0
Bit 6= 1
Bit 7= 0
Response MSaa aa is an ASCII character.
Errors D Execution not allowed.
S Communication time-out.
Example MS |Read acquisition status.
MS@ |System resting.
Related commands TB Read error.
TE Read error code.