EDH0170FE1010 – 09/98 48
LDS1000 Controller for Laser Diode Sensors
10.5.12 TA Query Sensor Parameters
Use h MNL j RMT j ACT
Syntax xxTA
Parameter xx Axis number: 1 for Y and 2 for Z.
Possible errors
xx ≠1 and xx ≠2 ☞ Error B.
Description Read parameters of a current sensor axis:
TYPE: RT or TR.
COEF: 00100 to 29999.
INCR: 0,1; 0,2; 0,5; 1; 2; 5; 10; 20; 50.
UNIT: µrad; sec; mrad; nm; µm; mm; unit.
OFS1: 00000 to 32767.
OFS2: 00000 to 32767.
The parameters are separated by commas.
Response Sensor axis parameters.
Errors B Incorrect axis number.
D Execution not allowed.
S Communication time-out.
Example 1TA |Read the parameters of axis 1 (Y) of the current sensor.
1TA,RT,04512,1,µrad,00102,00241 |Typical parameters for an STR500-H autocollimator.
Related commands TB Read error.
TE Read error code.
10.5.13 TB Error Description
Use h MNL j RMT j ACT
Syntax TBaa
Parameter aa Error code in ASCII format (@ to U).
If there is no parameter ☞ read current error.
Description Read message related to a an error code (parameter aa). Read
current error message if aa is missing.
Response TBaabb aa = Error code queried.
bb = Error description.
Errors D Execution not allowed.
S Communication time-out.
Example TB@ |Read error message @.
TB@ No error |No error.
Related command TE Read error code.