55 EDH0170FE1010 – 09/98
LDS1000 Controller for Laser Diode Sensors
10.5.22 VE Read Internal Program Version
Use h MNL j RMT j ACT
Syntax VE
Parameter None.
Description Read internal program version of the controller.
Response VE LDS1000 Controller Version xx.yy
xx = Version.
yy = Progress index.
Errors D Execution not allowed.
S Communication time-out.
Example VE |Read program version.
VE LDS1000 Controller Version 3.10 |Program version 3.10.
Related commands TB Read error.
TE Read error code.
10.5.23 XI Read Integration Constant
Use h MNL j RMT j ACT
Syntax XI
Parameter None.
Description Read integration constant given in milliseconds.
Response XInn nn is the integration constant given in milliseconds.
Errors D Execution not allowed.
S Communication time-out.
Example XI |Read integration constant.
XI100 |The integration constant is equal to 100 msec.
Related commands KI Define the integration constant.
TB Read error.
TE Read error code.