EDH0170FE1010 – 09/98 46
LDS1000 Controller for Laser Diode Sensors
10.5.09 RS Reset System
Use h MNL j RMT j ACT
Syntax RS
Parameter None.
Description Stop work in progress and carry out a RESET on the LDS1000
electronic controller.
Response None.
Error D Execution not allowed.
Example RS |Reset system.
Related commands TB Read error.
TE Read error code.
10.5.10 SP Define Sample Period
Use h MNL j RMT h ACT
Syntax SPnn
Parameter nn Sample period in milliseconds.
nn = 0.5 to 32000 msec, by steps of 0.5 msec.
Possible errors
nn absent nn = 0.5 msec.
nn off limits nn = Nearest authorized value.
Description Sets the sample period to nn milliseconds for TM and TK func-
tions.
Response None.
Error D Execution not allowed.
Example SP50 |Set the sample period to 50 msec.
Related commands TB Read error.
TE Read error code.
TK Acquisitions with trigger on TRIGG input.
TM Immediate acquisitions.