51 EDH0170FE1010 – 09/98
LDS1000 Controller for Laser Diode Sensors
10.5.16 TM Immediate Acquisition
Use h MNL j RMT h ACT
Syntax TMnn
Parameter nn 1 at NBMAX. Number of measurements to be performed
(NBMAX = Max. number of measurements related to the memo-
ry installed).
Error possible
nn outside authorized interval ā˜ž Error C
Description Immediately carries out an acquisition of nn successive measu-
rements.
The acquisition period is given by SP.
The integration constant set by KI is used.
The number of the acquisition in progress is given by XN.
XN is equal to nn when the series of acquisitions is finished.
Response None.
Errors C Parameter out of limit.
D Execution not allowed.
Example TM100 |Launch the acquisition of a series of 100 measurements.
Related commands KI Define the integration constant.
TB Read error.
TE Read error code.
SP Define sample period.
XN Read number of measurements in memory.
10.5.17 TN Read Number of Selected Sensor
Use h MNL j RMT j ACT
Syntax TN
Parameter None.
Description Read number of sensor selected by the user.
Response TNxx,yy xx is the number of the selected sensor (0 to 9).
yy is its series number.
Errors D Execution not allowed.
S Communication time-out.
Example TN |Request the number of the sensor selected.
TN5,151232 |Sensor number 5 is selected, its series number is 151232.
Related commands TB Read error.
TE Read error code.