EasyManuals Logo

Newport LDS1000 User Manual

Default Icon
77 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
47 EDH0170FE1010 – 09/98
LDS1000 Controller for Laser Diode Sensors
10.5.11 ST Stop Work in Progress
Use h MNL j RMT j ACT
Syntax ST
Parameter None.
Description Stop work in progress (acquisition launched by TK, TM or TQ).
Response None.
Error D Execution not allowed.
Example ST |Stop the work in progress and carry on.
NOTE
The ST command stops the acquisitions in progress without deleting the
memory. The measurements performed up to receiving this command
are stored in the controller’s memory.
The XN parameter keeps its value.
The command xxTT makes it possible to read the values stored if xx is
lower than XN.
Related commands TB Read error.
TE Read error code.
TK Acquisitions with trigger on TRIGG input.
TM Immediate acquisitions.
TQ Acquisitions with synchronization on TRIGG input.
TT Proof read acquisitions.
XN Read number of measurements in memory.