45 EDH0170FE1010 – 09/98
LDS1000 Controller for Laser Diode Sensors
10.5.07 RE Set Active Display Mode
Use h MNL j RMT j ACT
Syntax RE
Parameter None.
Description Enable display in remote mode. The response times related to
display refresh may appear.
Response None.
Error D Execution not allowed.
Example RE |Active display.
Related commands RD Frozen display.
TB Read error.
TE Read error code.
10.5.08 RO Read PWOK Output Status
Use h MNL j RMT j ACT
Syntax RO
Parameter None.
Description Read the status of the PWOK output. The return value will be
equal to 0 or 1 depending on whether the interface is in high or
low status.
Response ROnn nn = status of the PWOK output.
Errors D Execution not allowed.
S Communication time-out.
Example RO |Request status of TRIGG input.
RO0 |The PWOK output is in low status.
NOTE
The PWOK output is in status 1 if the sensor receives a sufficient quanti-
ty of light to carry out the measurement.
This logical output is available on the rear panel of the LDS1000 Controller.
Related commands TB Read error.
TE Read error code.