EasyManua.ls Logo

Newport LDS1000 - XN Read Number of Stored Measurements; XS Read Sample Period

Default Icon
77 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EDH0170FE1010 – 09/98 56
LDS1000 Controller for Laser Diode Sensors
10.5.24 XN Read Number of Stored Measurements
Use h MNL j RMT j ACT
Syntax XN
Parameter None.
Description Read number of measurements available in the memory.
Response XNnn The memory contains nn measurements.
Errors D Execution not allowed.
S Communication time-out.
Example XN |Read number of available measurements.
XN100 |The memory contains 100 measurements.
Related commands TB Read error.
TE Read error code.
TK Acquisitions with trigger on TRIGG input.
TM Immediate acquisitions.
TQ Acquisitions with synchronization on TRIGG input.
TT Re-read acquisitions.
10.5.25 XS Read Sample Period
Use h MNL j RMT j ACT
Syntax XS
Parameter None.
Description Read sample period.
Response XSnn The sample period is nn milliseconds.
Errors D Execution not allowed.
S Communication time-out.
Example XS |Read sample period.
XS2.5 |The sample period is equal to 2.5 ms (400 Hz).
Related commands TB Read error.
TE Read error code.
SP Define sample period.

Table of Contents