EDH0170FE1010 – 09/98 54
LDS1000 Controller for Laser Diode Sensors
10.5.21 TT Read Measurements Saved
Use h MNL j RMT j ACT
Syntax xxTT
Parameter xx Number of the acquisition to be proof read.
xx = 0 to MAX (MAX = value returned by XN).
Possible error
xx missing ☞ xx = 0.
xx outside authorized interval ☞ Error C.
Description This command enables the measurements saved to be read
again following a TK, TM or TQ command.
If xx is a whole number other than nil within the authorized
interval, the controller returns the value of the measurement
xx. If xx = 0 or absent, the controller returns all the measu-
rements in the memory (one measurement Y, Z per line star-
ting with the number 1).
In RS-232-C mode there is a 25-msec pause between two suc-
cessive lines (response at TT or 0TT) to avoid overflow.
Reading is possible as soon as the acquisitions have begun.
Response xxTT,1TPnn1,2TPnn2
Errors C Parameter out of limit.
D Execution not allowed.
S Communication time-out.
Example 9TT |Read measurement number 9.
9TT,1TP12.2,2TP258.1 |Measurement number 9: Y = 12.2 and Z = 258.1.
Related commands TB Read error.
TE Read error code.