Causes and countermeasures
condition, check the connections or replace the
cuff, and then re-test.
The used cuff does not match the set patient
category. Verify the patient category and replace
the cuff.
NIBP cuff isn’t placed or connected properly, or
there is gas leak.
Check cuff and inflation tube.
Ambient atmospheric pressure is abnormal.
Confirm that the environment complies with the
monitor's specifications, and check whether there
are special reasons affecting ambient pressure.
The measured blood pressure of the patient
exceeds the measuring range.
Patient's pulse may be weak or cuff is too loose.
Check the condition of the patient, and place the
cuff in a suitable position. If the failure persists,
replace the cuff.
Excessive movement may result in too much
motion artifact or interference in the signal during
measurement.
Motion signal amplitude is too large due to
movement and other reasons.
Cuff overpressure, and gas blockage may occur;
check the gas path, and then re-measure.
NIBP module reset error; check the gas path is
blocked, and then restart the measurement.